Studies of self assembled monolayers on highly oriented pyrolytic graphite using scanning tunneling microscopy and computational simulation 3

14 213 0
Studies of self assembled monolayers on highly oriented pyrolytic graphite using scanning tunneling microscopy and computational simulation 3

Đang tải... (xem toàn văn)

Tài liệu hạn chế xem trước, để xem đầy đủ mời bạn chọn Tải xuống

Thông tin tài liệu

CALIBRATION CHAPTER CALIBRATION OF SCANNING TUNNELING MICROSCOPY USING HIGHLY ORIENTED PYROLYTIC GRAPHITE AS STANDARD 3.1 Purpose of Calibration The scanning tunneling microscope (STM) is the ideal instrument for surface investigations of matter on an atomic scale. The most quantitative information obtained by STM is the geometry and physics of surface topography, including the roughness of microstructures [1, 2]. Incorrect lateral calibration of the STM scanning elements will cause the images to be stretched, compressed or even skewed. Thus it is crucial to have accurate calibration to achieve reliable STM images. Several factors can cause the distortion of the STM images, including temperature and humidity which affect the piezoceramic’s sensitivity, hysteresis of the system, thermal drift, the sharpness of the STM tip, and noise. In order to minimize these effects, a very sharp Pt/Ir tip was used throughout scanning process. The vibration isolation system is in operation during the experiment to isolate the vibration from the environment. The experimental room is equipped with air-conditioners and dehumidifier to control the temperature and humidity respectively. Some works have been carried out regarding the calibration of STM. Jorgensen and coworkers [3] developed a technique based on detection of the reciprocal unit cell in Fourier space for estimating the lateral calibration factors and the drift in the fast scanning direction. Desogus and coworkers used a capacitive sensor coupled to a 38 CALIBRATION STM piezoscanner for accurate measurements [4]. Lapshin also described a practical method to find automatically the calibration coefficients and residual nonorthogonality of STM [5]. Wang et al employed the combination of STM and X-Ray interferometer for calibration [2]. There are also some computational methods developed for data analysis to remove the distortion [6, 7]. However, these methods are not suitable for our experiments, either due to their complexity or incompatibility with our instruments. Here we present a simple calibration procedure by improving the original calibration method developed by Veeco Instruments. 3.2 Piezoelectricity and Scanner Piezoelectricity is the ability of some materials (deleted) to generate an electric potential in response to applied mechanical stress. This may take the form of a separation of electric charges across the crystal lattice. The piezoelectric effect is reversible, exhibiting the direct piezoelectric effect (the production of electricity when stress is applied) as well as the converse piezoelectric effect (the production of stress and/or strain when an electric field is applied). Lead zirconate titanate (PZT) crystal is a ceramic perovskite material that shows a marked piezoelectric effect [8]. The effect finds useful applications such as the production and detection of sound, generation of high voltages, electronic frequency generation, microbalance, and ultra fine focusing of optical assemblies. It is also the basis of operation of the scanning probe microscope, such as STM and AFM. Scanner serve as the “movers and shakers” of every scanning probe microscope. 39 CALIBRATION Typically, they consist of a hollow tube made of piezoelectric material such as PZT [9]. Piezo materials contract or elongate when a negative or positive voltage is applied depending upon the orientation of the material’s polarized grain structure. The scanners are used to manipulate the sample-tip movement very precisely to scan the sample surface. In some models (e.g., Small Sample MultiMode) the scanner tube moves the sample, with the tip set stationary. In other models (e.g., Dimension SPMs) the sample is stationary and the tip is moved. The different scanners not react exactly the same to a voltage. Because of slight variations in the orientation and size of the piezoelectric granular structure (polarity), material thickness, etc., each scanner has a unique property [9]. This property is conveniently measured in terms of sensitivity, a ratio of piezo voltage-to-piezo movement. Sensitivity is not a linear relationship, because piezo scanners exhibit a larger sensitivity (i.e., more movement per volt) at higher voltages than they at lower voltages. This non-linear relationship is determined for each scanner crystal. As the scanner ages, its sensitivity will decrease and periodic recalibration becomes necessary. The response curve of piezo-materials to applied voltage curve is presented in Fig 3.1. This curve typifies scanner sensitivity across the full range of movement. The vertical axis denotes the voltage applied to the scanner [9] while the horizontal axis denotes the scanner movement. At higher voltages, the scanner’s sensitivity increases (i.e. more movement per voltage applied), whereas the scanner is “motionless” at zero volts. Plotting each point along the curve describes a second-order, exponential 40 CALIBRATION relationship which provides a rough approximation of the scanner sensitivity. Fig 3.1 Diagram showing the response curve of piezo-materials to applied voltage curve [9]. The vertical axis denotes the voltage applied to the scanner while the horizontal axis denotes the scanner movement. The scanner’s sensitivity is not linear, with larger sensitivity at higher voltage. 3.3 Calibration of STM 3.3.1 Calibration Methods For purposes of calibration, the manufacturer employs various derating and coupling parameters to model the scanners’ nonlinear characteristics. By precisely determining the points along the scanner’s sensitivity curve, and through the application of a rigorous mathematical model, the full-range measuring accuracy can be made (deleted) better than percent. Because the scanner sensitivities vary according to the voltage applied to them, the reference must be thoroughly scanned at a variety of sizes and angles. The user then specifies the distance between known features on the reference’s surface and a parameter is recorded to compensate the scanner’s movements The X, Y and Z axes may be calibrated in any sequential order. 41 CALIBRATION The calibration procedure consists of three steps. Firstly, a calibration reference having surface features of precisely known dimensions is scanned. These dimensions are compared with those estimated by the SPM software; and the calibration parameters are adjusted until the SPM’s dimensions meet the true dimensions of the reference. The “A” scanner is the smallest scanner, with a total travel distance of approximately 0.4m along each axis. Its compact design provides excellent stability for atomic scans. Based on the previous method developed by the manufacturer, we developed a relatively simple methodology to perform calibration of the “A” scanner for scanning tunneling microscope function. 3.3.2 Calibration Procedures The calibration of STM can be conducted with a standard sample: highly ordered pyrolytic graphite (HOPG). The HOPG needs to be cleaved prior to the scanning. Cleaving was accomplished by the adhering of a tape to the surface and then having it peeled off, which produced a fresh surface of atoms having a regular lattice. The sample to be used was placed on a puck and then attached to the scanner cap. The surface was engaged by the tip. The Intergral Gain and Setpoint are adjusted to obtain good images. The scanner was allowed to stabilize by scanning 2-3 frames. If the scanner has been running at extreme voltages, the stabilization time might take longer. The Setpoint is kept low if possible. In our calibration, its value is set between 30-50pA. The Z Center Position is set to 0V. The scan rate is set quite high (~12Hz) 42 CALIBRATION for atomic scale images to reduce some of the noise due to thermal drift. The condition of the tip is critical for obtaining a good STM image with atomic resolution. Fig 3.2 is a typical STM image of HOPG scanned with a poorly prepared Pt-Ir tip. Fig 3.2 STM image of a cleaved HOPG scanned with a poorly prepared Pt-Ir tip prior to calibration (6×6 nm, Vbias=80mV, Ioffset=30pA) If difficulty is experienced obtaining an image, a different Pt-Ir tip will be installed until the image of carbon atoms can be roughly observed. Fig 3.3 shows a STM image of HOPG scanned with a good Pt-Ir tip. 43 CALIBRATION Fig 3.3 Typical STM image a clean HOPG scanned with a good Pt-Ir tip prior to calibration (6×6 nm, Vbias=80mV, Ioffset=30pA). Image was treated with Lowpass function of the Nanoscope IIIa software to remove noise. The alignment of the graphite crystal on the working platform was not always ideal; it resulted in the angle between the measured direction and horizontal direction. Therefore the line containing the highest number of bright carbon atoms was selected for distance measurement. The section analysis (Fig 3.4) of the distance profile along the white arrow in Fig 3.3 showed that the average distance between neighbouring bright carbon atoms was 0.267nm. 44 CALIBRATION Fig 3.4 Section Analysis of the STM image in Fig 3.3 along the orientation of the white arrow Fig 3.5 shows the standard parameters of the graphite surface lattice, where the real atomic spacing between neighbouring bright carbon atoms is 0.246nm. The difference between the measured distance and the literature value was 8.5%, which was more than the allowed error of 2.0%. Thus, the calibration procedure had to be carried out to ensure the measuring error of the STM measurements fall within the allowed value. Fig 3.5 Schematic diagram of atomic spacing on the HOPG surface under STM. Each circle represents one bright carbon atoms. 45 CALIBRATION The calibration parameters were listed in Fig 3.6. Among them the ‘X fast sens’, ‘Y fast sens’, ‘X slow sens’ and ‘Y slow sens’ should be modified accordingly. We started with “X fast sens” when the scan angle was set at 0°. The new “X fast sens” value theoretical distance  original value  fast sens value measured distance 0.246 =  3.100  2.814 0.271 = The new values of “Y slow sens” (at 0°, vertical direction), “Y fast sens” (at 90°, horizontal direction) and “X slow sens” (at 90°, vertical direction) were also calculated. After four new parameters were employed, the STM images of HOPG were captured again after scanning for a certain period of time when whole work station became stable. Fig 3.6 Calibration parameters of scanner A The new value of spacing between the adjacent bright carbons was recalculated. 46 CALIBRATION If the difference between the experimental result and the literature value was still larger than 2%, a re-calibration was necessary. Fig 3.7 shows the final graphite image obtained by the STM after repeating several times of calibration procedures. The measured distance between adjacent bright carbon atoms was 0.250nm, as shown by the section analysis, which was within the allowed range of 0.241 to 0.251nm. The calibration was considered successful and subsequent experiment would be carried out using the same set of parameters obtained based on the latest calibration. Fig 3.7 STM image of HOPG after calibration with a good Pt-Ir tip (6×6nm, Vbias=80mV, Ioffset=30pA). Image was treated with Lowpass function of the Nanoscope IIIa software to remove noise. 47 CALIBRATION Fig 3.8 Section Analysis of the STM image in Fig 3.7 3.4 Discussion The calibration of the STM is particularly important in our experiment and a chapter is therefore dedicated to this subject. The calibration method provided by the manufacturer has been modified. For example, for the parameter of “X fast sens”, we are supposed to measure the distance horizontally. However, in most calibration procedures it is difficult to position the graphite crystal at the correct angle to carry out the measurements. Instead, the distance along the line of highest atom density white arrow shown in Fig 3.3 was measured. The calibration can be easily done by repeating that method several times, without spending time on positioning the graphite crystal on the scanner platform. During the scanning process, the tip moved in two directions: from the top of the image to the bottom of the image, and in the opposite direction. It was observed that 48 CALIBRATION the images obtained when the tip moves in the opposite directions were different. This could be due to the asymmetric form of the STM tip. The Pt/Ir tip was usually cut mechanically along the 45° angle, and therefore the tip did not have a symmetric shape. In addition the STM tip was not perpendicular to the scanning surface in most cases, as shown in Fig 3.9. Fig 3.9 Illustration of contact between the STM tip and the HOPG surface As the STM images were scanning-direction dependent, only images at one scanning direction were collected to ensure the consistency of the experimental results. The STM images can also be affected by the height of the sample. When the height of the sample was changed, the STM head had to be adjusted, which in turn changed the orientation of Pt/Ir tip and caused distortion of the images. It is important to keep the sample height constant during the whole experiments. Hence, the same HOPG crystal was used throughout the experiment, including calibration, SAMs preparation and STM measurements. Furthermore, the piezoceramic component is very sensitive to the environment, 49 CALIBRATION such as temperature and humidity. The working environment has been equipped with air conditioner and dehumidifier to provide the constant temperature and humidity for STM instruments. It was also strongly recommend that all the data should be collected on the same day for one sample. 3.5 Conclusion In short, by modifying the original calibration methods provided by the manufacturer, we developed the calibration procedures which are more convenient for users. Due to the complexity and sensitivity of the STM components, calibration must be carried out every time before studying of the surfaces to ensure the reliability of the experimental results. 50 CALIBRATION References [1] Pope, K.J.; Smith, J.L.; Shapter, J.G. Smart Mater. Struct. 2002, 11, 679. [2] Lin, W.; Kuetgens, U.; Becker, P.; Koenders, L.; Li, D.C.; Cao, M. Nanotechnology 1999, 10, 412. [3] Jorgensen, J.F.; Madsen, L.L.; Garnaes, J.; Carneiro, K.; Schaumburg, K. J. Vac. Sci. Technol. B 1994, 12, 1698. [4] Desogus, S.; Lanyi, S.; Nerino, R.; Picotto, G.B. J. Vac. Sci. Technol. B 1994, 12, 1665. [5] Lapshin, R.V. Rev. Sci. Instrum., 1998, 69, 3268. [6] Andersen, J. E. T.; Møller, P. Surf. Coat. Technol. 1994, 67, 213. [7] Lapshin, R.V. Meas. Sci. Technol. 2007, 18, 907. [8] Gandhi, M.V.; Thompson, B.S. Smart Materials and Structures, London: Chapman & Hall, 1992, 1st edition. [9] Digital Instruments, Support Note No. 217, Rev. D 51 [...]... perpendicular to the scanning surface in most cases, as shown in Fig 3. 9 Fig 3. 9 Illustration of contact between the STM tip and the HOPG surface As the STM images were scanning- direction dependent, only images at one scanning direction were collected to ensure the consistency of the experimental results The STM images can also be affected by the height of the sample When the height of the sample was... working environment has been equipped with air conditioner and dehumidifier to provide the constant temperature and humidity for STM instruments It was also strongly recommend that all the data should be collected on the same day for one sample 3. 5 Conclusion In short, by modifying the original calibration methods provided by the manufacturer, we developed the calibration procedures which are more convenient...CALIBRATION Fig 3. 8 Section Analysis of the STM image in Fig 3. 7 3. 4 Discussion The calibration of the STM is particularly important in our experiment and a chapter is therefore dedicated to this subject The calibration method provided by the manufacturer has been modified For example, for the parameter of “X fast sens”, we are supposed to measure the distance horizontally However, in most calibration procedures... position the graphite crystal at the correct angle to carry out the measurements Instead, the distance along the line of highest atom density white arrow shown in Fig 3. 3 was measured The calibration can be easily done by repeating that method several times, without spending time on positioning the graphite crystal on the scanner platform During the scanning process, the tip moved in two directions:... changed the orientation of Pt/Ir tip and caused distortion of the images It is important to keep the sample height constant during the whole experiments Hence, the same HOPG crystal was used throughout the experiment, including calibration, SAMs preparation and STM measurements Furthermore, the piezoceramic component is very sensitive to the environment, 49 CALIBRATION such as temperature and humidity The... directions: from the top of the image to the bottom of the image, and in the opposite direction It was observed that 48 CALIBRATION the images obtained when the tip moves in the opposite directions were different This could be due to the asymmetric form of the STM tip The Pt/Ir tip was usually cut mechanically along the 45° angle, and therefore the tip did not have a symmetric shape In addition the STM tip was... Picotto, G.B J Vac Sci Technol B 1994, 12, 1665 [5] Lapshin, R.V Rev Sci Instrum., 1998, 69, 32 68 [6] Andersen, J E T.; Møller, P Surf Coat Technol 1994, 67, 2 13 [7] Lapshin, R.V Meas Sci Technol 2007, 18, 907 [8] Gandhi, M.V.; Thompson, B.S Smart Materials and Structures, London: Chapman & Hall, 1992, 1st edition [9] Digital Instruments, Support Note No 217, Rev D 51 ... to the complexity and sensitivity of the STM components, calibration must be carried out every time before studying of the surfaces to ensure the reliability of the experimental results 50 CALIBRATION References [1] Pope, K.J.; Smith, J.L.; Shapter, J.G Smart Mater Struct 2002, 11, 679 [2] Lin, W.; Kuetgens, U.; Becker, P.; Koenders, L.; Li, D.C.; Cao, M Nanotechnology 1999, 10, 412 [3] Jorgensen, J.F.; . CALIBRATION CHAPTER 3 CALIBRATION OF SCANNING TUNNELING MICROSCOPY USING HIGHLY ORIENTED PYROLYTIC GRAPHITE AS STANDARD 3. 1 Purpose of Calibration The scanning tunneling microscope. calibration of the “A” scanner for scanning tunneling microscope function. 3. 3.2 Calibration Procedures The calibration of STM can be conducted with a standard sample: highly ordered pyrolytic graphite. finds useful applications such as the production and detection of sound, generation of high voltages, electronic frequency generation, microbalance, and ultra fine focusing of optical assemblies.

Ngày đăng: 14/09/2015, 08:37

Từ khóa liên quan

Tài liệu cùng người dùng

  • Đang cập nhật ...

Tài liệu liên quan